CAM 100 Panel meter (48 x 24) |
 |
- DC-signals 0 to 200 V resp. 0 to 200 mA
- Temperature measurement with Pt 100-sensor
- 3½-digit 7 segment LED, red
- Display range from - 1999 to + 1999 free scaling
Display
- Display 7-Segment display, 8 mm LED red
- Display range ± 1999 digit
- Decimal point Via solder pads
- Error indication Display „1“ at overrange or sensor break
- Display „-1“ at underrange
|
 |
 |
 |
Features |
 |
- Meas. range attitude
Via solder pads and scaling resistor Rs at terminal
- Method of measurement
Dual slope
- Sampling rate
Approx. 2,5 / sec.
- Response time
< 2 sec. at 100% step
- Input signals
Voltage, current (DC), Pt 100
- Measuring range U/I
0 ... 10 V, 0 (4) ... 20 mA
± 0,2 / ± 2 / ± 20 / ± 200 V / mA
Accuracy: £ ± 1 digit ± 0,15 % of display range
Temp. effect: £ ± 0,08 digit / K (related to 25 °C)
Gain adjustment: max. ± 20 %
Offset adjustment: max. ± 500 digit
- Pt 100
- 100,0 ... + 199,9 °C
Accuracy: £ ± 0,2 K ± 0,2 % of display range
Temp. effect: £ ± 0,02 K / K (related to 25 °C)
- 200 ... + 700 °C
Accuracy: £ ± 1 K ± 0,2 % of display range
Temp. effect: £ ± 0,04 K / K (related to 25 °C)
Connection for 2-wire (max. 3 W resistance), 3- or 4-wire sensors
(each. max. 100 W resistance)
Wire effect: 3-wire connection: £ ± 0,01 K / W
4-wire connection: £ ± 0,002 K / W
Distribution: 4-wire connection, meas. range: -100,0 ... +199,9°C
|
 |
Delivery:
|
 |
| CAM 100 Panel meter (48 x 24) |
 |
|